Vascular Complications in Transcatheter Aortic Valve Replacement With Plug-Based vs Suture-Based Closure Devices

Lior Zornitzki*, David Zahler, Shir Frydman, Tamar Itach, Jeremy Ben-Shoshan, Amir Halkin, Samuel Bazan, Yaron Arbel, Maayan Konigstein, Ariel Finkelstein, Shmuel Banai, Arie Steinvil

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Background: There are conflicting data regarding the efficacy and safety of suture vs plug-based vascular closure devices (VCDs) for large-bore catheter management in patients undergoing transcatheter aortic valve replacement (TAVR). We compared the rates of vascular complications (VCs) associated with 2 commonly used VCDs in a large cohort of patients undergoing TAVR. Methods: We conducted a single-centre, all-comer, prospective registry study, enrolling patients undergoing TAVR for symptomatic severe aortic stenosis (AS) between the years 2009 and 2022. Clinical outcomes were compared between patients undergoing closure of the femoral access point using the MANTA VCD (M-VCD) (Teleflex, Wayne, PA) vs the ProGlide VCD (P-VCD) (Abbott Vascular, Abbott Park, IL). The main outcome measures were researcher adjudicated events of VARC-2 defined major and minor VCs. Results: Overall, 2368 patients were enrolled in the registry; 1315 (51.0% male, 81.0 ± 7.0 years) patients were included in the current analysis. P-VCD was used in 813 patients, whereas M-VCD was used in 502 patients. In-hospital VCs were more frequent in the M-VCD vs the P-VCD group (17.3% vs 9.8%; P < 0.001). This outcome was mainly driven by elevated rates of minor VCs in the M-VCD group, whereas no significant difference was observed for major VCs (15.1% vs 8.4%; P < 0.001 and 2.2% vs 1.5%; P = 0.33, respectively). Conclusions: In patients undergoing TAVR for severe AS, M-VCD was associated with higher rates of VCs. This outcome was mainly driven by minor VCs. The rate of major VCs was low in both groups.

Original languageEnglish
Pages (from-to)1528-1534
Number of pages7
JournalCanadian Journal of Cardiology
Volume39
Issue number11
DOIs
StatePublished - Nov 2023

Fingerprint

Dive into the research topics of 'Vascular Complications in Transcatheter Aortic Valve Replacement With Plug-Based vs Suture-Based Closure Devices'. Together they form a unique fingerprint.

Cite this