Some low temperature thermometry observations

Research output: Contribution to journalArticlepeer-review

Abstract

The temperature calibration on annealed Chromel-P/Au single bond Fe thermocouples has been extended from 1.3 to 0.55 K. Temperature calibration tables on annealed copper/Au single bond Fe thermocouples are presented. Electromotive force constants representing the potentials from annealed Au single bond Fe thermocouples between liquid helium and liquid nitrogen temperature references and between liquid helium and ice temperature references have been measured. Through the use of these emf constants, the calibration tables have been shifted and may be used either with an ice bath or liquid nitrogen temperature reference bath. Temperature calibrations on small platinum sensors manufactured by Minco Products, Inc. are given.

Original languageEnglish
Pages (from-to)577-583
Number of pages7
JournalReview of Scientific Instruments
Volume40
Issue number4
DOIs
StatePublished - 1969
Externally publishedYes