T1 - Potential dosing errors using portable prothrombin time monitoring devices
AU - Loebstein, Ronen
AU - Kurnik, Daniel
AU - Lubetsky, Aharon
AU - Ezra, David
AU - Halkin, Hillel
PY - 2003/7
Y1 - 2003/7
N2 - Portable prothrombin time (PT) monitors facilitate the control of warfarin therapy. Few studies have compared the influence of using different monitors on dosage decisions. We determined the comparability of data generated by two portable PT monitors, Coaguchek S, (Roche Diagnostics Boehringer-Mannheim) and Hemochron Jr (International Technidyne Corporation Ltd.), with that of a reference laboratory. Simultaneous International Normalized Ratio (INR) measurements (portable monitor and laboratory) were performed in 193 consecutive patients receiving warfarin for at least 3 months. Agreement of measurements was assessed by both regression analysis and influence on dosage decisions in accordance with pre-defined criteria. The Coaguchek S versus laboratory INR regression line (n = 111; r2 = 0.88; P < 0.001) was close to the line of identity, while that of the Hemochron Jr (n = 82; r2 = 0.61; P < 0.001) was not. The overall proportion of dual INR measurements that fulfilled the clinical criteria of agreement was 90% for the Coaguchek S compared with 62% for the Hemochron Jr (P < 0.0001). For laboratory INRs 2.0-2.5, 2.6-4.0 and > 4.0, the proportions of portable measurements that satisfied the clinical criteria for the Coaguchek S versus the Hemochron Jr were 96 versus 63% (P < 0.001), 81 versus 45% (P < 0.04), and 67 versus 17% (P < 0.85), respectively. Warfarin dosing based solely on the portable devices would have resulted in unjustified dose increments in 22% of the patients with the Hemochron Jr device compared with 8% with the Coaguchek S monitor (χ2 = 4.43; P = 0.035). The Coaguchek S monitor provides measurements for INR values within the therapeutic range that agree well with the standard laboratory. The Hemochron Jr measurements result in different dosage adjustments even within the therapeutic range, but especially for INR values > 4.0. For both monitors, agreement of INR measurements with the standard decreases with increasing INR values.
AB - Portable prothrombin time (PT) monitors facilitate the control of warfarin therapy. Few studies have compared the influence of using different monitors on dosage decisions. We determined the comparability of data generated by two portable PT monitors, Coaguchek S, (Roche Diagnostics Boehringer-Mannheim) and Hemochron Jr (International Technidyne Corporation Ltd.), with that of a reference laboratory. Simultaneous International Normalized Ratio (INR) measurements (portable monitor and laboratory) were performed in 193 consecutive patients receiving warfarin for at least 3 months. Agreement of measurements was assessed by both regression analysis and influence on dosage decisions in accordance with pre-defined criteria. The Coaguchek S versus laboratory INR regression line (n = 111; r2 = 0.88; P < 0.001) was close to the line of identity, while that of the Hemochron Jr (n = 82; r2 = 0.61; P < 0.001) was not. The overall proportion of dual INR measurements that fulfilled the clinical criteria of agreement was 90% for the Coaguchek S compared with 62% for the Hemochron Jr (P < 0.0001). For laboratory INRs 2.0-2.5, 2.6-4.0 and > 4.0, the proportions of portable measurements that satisfied the clinical criteria for the Coaguchek S versus the Hemochron Jr were 96 versus 63% (P < 0.001), 81 versus 45% (P < 0.04), and 67 versus 17% (P < 0.85), respectively. Warfarin dosing based solely on the portable devices would have resulted in unjustified dose increments in 22% of the patients with the Hemochron Jr device compared with 8% with the Coaguchek S monitor (χ2 = 4.43; P = 0.035). The Coaguchek S monitor provides measurements for INR values within the therapeutic range that agree well with the standard laboratory. The Hemochron Jr measurements result in different dosage adjustments even within the therapeutic range, but especially for INR values > 4.0. For both monitors, agreement of INR measurements with the standard decreases with increasing INR values.
KW - Accuracy
KW - Clinical agreement
KW - Portable prothrombin time monitors
UR - http://www.scopus.com/inward/record.url?scp=0042626436&partnerID=8YFLogxK
U2 - 10.1097/00001721-200307000-00008
DO - 10.1097/00001721-200307000-00008
M3 - ???researchoutput.researchoutputtypes.contributiontojournal.article???
AN - SCOPUS:0042626436
SN - 0957-5235
VL - 14
SP - 479
EP - 483
JO - Blood Coagulation and Fibrinolysis
JF - Blood Coagulation and Fibrinolysis
IS - 5
ER -