Pneumothorax: An unusual cause of ICD defibrillation failure

David Luria, Marshal S. Stanton, Michael Eldar, Michael Glikson*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

17 Scopus citations

Abstract

We describe two patients with defibrillation failure of implantable cardioverter defibrillators (ICDs) resulting from large left pneumothoraxes following subclavian vein puncture during the implantation. Following pneumothorax drainage, low defibrillation thresholds (DFTs) were attained without further manipulations. The absence of other signs and symptoms of pneumothorax and the presence of satisfactory pacing function during the procedure, resulted in a significant delay in diagnosis. Pneumothorax should be included in the differential diagnosis when unexpected high DFTs are found during ICD implantation or predischarge testing. This complication is avoidable by a different surgical approach, cephalic vein cutdown.

Original languageEnglish
Pages (from-to)474-475
Number of pages2
JournalPACE - Pacing and Clinical Electrophysiology
Volume21
Issue number2
DOIs
StatePublished - 1998

Keywords

  • Defibrillation threshold
  • Implantable cardioverter defibrillator
  • Pneumothorax