Monitoring lung resistivity changes in congestive heart failure patients using the bioimpedance technique.

Sharon Zlochiver*, Michal M. Radai, Deganit Barak-Shinar, Tuiva Ben-Gal, Vicky Yaari, Boris Strasberg, Shimon Abboud

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

15 Scopus citations

Abstract

The feasibility of a novel, dedicated system for monitoring lung resistivity in congestive heart failure patients, implementing a hybrid approach of the bioimpedance technique, was assessed in this preliminary study. Thirty-three healthy volunteers and 34 congestive heart failure patients were measured with the PulmoTrace system (CardioInspect, Tel Aviv University, Tel Aviv, Israel) during tidal respiration, and the ability to monitor the respective lung resistivity values was assessed. Mean left and right lung resistivity values of 1205+/-163 and 1200+/-165 ohm.cm for the control group and 888+/-193 and 943+/-187 ohm.cm for the congestive heart failure group were found, indicating a significant (p<2.10(-7)) difference between the two groups. The results of long-term monitoring of two patients during medical treatment are also shown. This hybrid approach system is believed to improve diagnostic capabilities and help physicians to better adjust medication dosage on a frequent basis.

Original languageEnglish
Pages (from-to)289-293
Number of pages5
JournalPrevention and Management of Congestive Heart Failure
Volume11
Issue number6
DOIs
StatePublished - 2005

Fingerprint

Dive into the research topics of 'Monitoring lung resistivity changes in congestive heart failure patients using the bioimpedance technique.'. Together they form a unique fingerprint.