In-Clinic Evaluation of the MiniMed 670G System "suspend before Low" Feature in Children with Type 1 Diabetes

Michael A. Wood*, Dorothy I. Shulman, Gregory P. Forlenza, Bruce W. Bode, Orit Pinhas-Hamiel, Bruce A. Buckingham, Kevin B. Kaiserman, David R. Liljenquist, Timothy S. Bailey, John Shin, Suiying Huang, Xiaoxiao Chen, Toni L. Cordero, Scott W. Lee, Francine R. Kaufman

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

Background: The Medtronic predictive low-glucose management (PLGM) algorithm automatically stops insulin delivery when sensor glucose (SG) is predicted to reach or fall below a preset low-glucose value within the next 30 min, and resumes delivery after hypoglycemia recovery. The present study evaluated the PLGM algorithm performance of the MiniMed™ 670G system SmartGuard™ "suspend before low" feature in children aged 7-13 years with type 1 diabetes (T1D). Method: Participants (N = 105, mean ± standard deviation of 10.8 ± 1.8 years) underwent an overnight in-clinic evaluation of the "suspend before low" feature with a preset low limit of 65 mg/dL. After exercise, frequent sample testing (FST) was conducted every 5 min if values were <70 mg/dL; every 15 min if 70-80 mg/dL; and every 30 min if >80 mg/dL. First-day performance of the Guardian™ Sensor 3 glucose sensor and continuous glucose monitoring system was also evaluated. Results: Activation of the "suspend before low" feature occurred in 79 of the 105 participants, 79.7% (63/79) did not result in SG falling below 65 mg/dL. Mean glucose at activation was 102 ± 19 mg/dL and the initial insulin suspension duration was 87.5 ± 32.7 min. Four hours after insulin resumption, mean reference glucose was 130 ± 42 mg/dL. Mean absolute relative difference between the FST reference glucose and SG values on the first day of sensor wear was 11.4%. For the 26 participants in whom the "suspend before low" feature did not activate, none involved a reference glucose value ≤65 mg/dL, suggesting that the PLGM algorithm performed as intended. Conclusion: In children aged 7-13 years with T1D, the "suspend before low" feature of the MiniMed 670G system demonstrated a hypoglycemia prevention rate of nearly 80% after exercise and did not involve rebound hyperglycemia. There were no events of severe hypoglycemia during the evaluation.

Original languageEnglish
Pages (from-to)731-737
Number of pages7
JournalDiabetes Technology and Therapeutics
Volume20
Issue number11
DOIs
StatePublished - Nov 2018
Externally publishedYes

Keywords

  • Automated insulin delivery
  • Hybrid closed loop
  • Hypoglycemia
  • Insulin pump suspension
  • Pediatrics
  • Predictive low glucose management
  • Type 1 diabetes