Implications of hydrodynamic testing to guide sizing of self-expanding transcatheter heart valves for valve-in-valve procedures

Janarthanan Sathananthan, Mark Hensey, Rob Fraser, Uri Landes, Philipp Blanke, Hoda Hatoum, Lakshmi Prasad Dasi, Alexander Sedaghat, Vinayak N. Bapat, Jonathon Leipsic, Lars Søndergaard, David A. Wood, John G. Webb*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

Aims: The commonly used valve-in-valve (VIV) app recommends sizing based on dimensions of both the transcatheter heart valve (THV) and bioprosthetic surgical valve. The implications of hydrodynamic testing to guide VIV sizing are poorly understood. This bench study assessed the hydrodynamic performance of different sizes of self-expanding supra-annular THVs in three different surgical aortic bioprostheses at different implantation depths. Methods: A small versus medium ACURATE neo (ACn), and a 26 mm versus 29 mm Evolut R were assessed after VIV implantation in 25 mm Mitroflow, Mosaic, and Magna Ease aortic surgical bioprostheses, at three implantation depths (+2 mm, -2 mm, and -6 mm). Results: The medium-sized ACn had lower gradients compared to the small ACn when the THV was implanted high (+2 mm, or -2 mm). The 29 mm Evolut R had lower gradients compared to a 26 mm Evolut R for all implantation depths, except for a depth of -2 mm in the 25 mm Mitroflow. The medium ACn and 29 mm Evolut R had larger effective orifice areas compared to the small ACn and 26 mm Evolut R, respectively. Both Evolut R sizes had acceptable regurgitant fractions (<15%), while both ACn sizes were above the acceptable performance criteria (>15%), at all implantation depths. Conclusions: Use of a larger self-expanding THV was associated with superior hydrodynamic performance if the THV was implanted high. Hydrodynamic testing can provide additional information to the VIV app to help guide VIV sizing.

Original languageEnglish
Pages (from-to)E332-E340
JournalCatheterization and Cardiovascular Interventions
Volume96
Issue number3
DOIs
StatePublished - 1 Sep 2020
Externally publishedYes

Funding

FundersFunder number
CT core lab services for Edwards Lifesciences
ViVitro Labs
Abbott Laboratories
Medtronic
Edwards Lifesciences
Boston Scientific Corporation
Abbott Vascular
Symetis

    Keywords

    • ACURATE neo
    • hydrodynamic performance
    • self-expanding
    • transcatheter heart valve
    • valve-in-valve

    Fingerprint

    Dive into the research topics of 'Implications of hydrodynamic testing to guide sizing of self-expanding transcatheter heart valves for valve-in-valve procedures'. Together they form a unique fingerprint.

    Cite this