TY - JOUR
T1 - Coronary Sinus Reducer Stent for the Treatment of Chronic Refractory Angina Pectoris. A Prospective, Open-Label, Multicenter, Safety Feasibility First-in-Man Study
AU - Banai, Shmuel
AU - Ben Muvhar, Shmuel
AU - Parikh, Keyur H.
AU - Medina, Aharon
AU - Sievert, Horst
AU - Seth, Ashok
AU - Tsehori, Jonathan
AU - Paz, Yoav
AU - Sheinfeld, Ami
AU - Keren, Gad
PY - 2007/5/1
Y1 - 2007/5/1
N2 - Objectives: This study sought to evaluate the safety of the Coronary Sinus Reducer (Neovasc Medical, Inc., Or Yehuda, Israel) as a potential alternate therapy for patients with refractory angina who are not candidates for conventional revascularization procedures. Background: Increased coronary sinus (CS) pressure can reduce myocardial ischemia by redistribution of blood from nonischemic to ischemic territories. The Coronary Sinus Reducer is a percutaneous implantable device designed to establish CS narrowing and to elevate CS pressure. In preclinical experiments, implantation of the Reducer was safe and was associated with improved ischemic parameters. In the present study, the safety and feasibility of the Coronary Sinus Reducer was evaluated in patients with refractory angina who were not candidates for revascularization. Methods: Fifteen coronary artery disease patients with severe angina and reversible ischemia were electively treated with the Reducer. Clinical evaluation, dobutamine echocardiography, thallium single-photon emission computed tomography, and administration of an angina questionnaire were performed before and 6 months after implantation. Cardiac computed tomography was performed 2 days and 6 months after implantation. Results: All procedures were completed successfully. No procedure-related adverse events occurred during the periprocedural and the follow-up periods. Angina score improved in 12 of 14 patients. Average Canadian Cardiovascular Society score was 3.07 at baseline and 1.64 at follow-up (n = 14, p < 0.0001). Stress-induced ST-segment depression was reduced in 6 of 9 patients and was eliminated in 2 of these 6 (p = 0.047). The extent and severity of myocardial ischemia by dobutamine echocardiography and by thallium single-photon emission computed tomography was reduced (p = 0.004 [n = 13] and p = 0.042 [n = 10], respectively). Conclusions: Implantation of the Coronary Sinus Reducer is feasible and safe. These findings, along with the clinical improvement observed, support further evaluation of the Reducer as an alternative treatment for patients with chronic refractory angina who are not candidates for coronary revascularization.
AB - Objectives: This study sought to evaluate the safety of the Coronary Sinus Reducer (Neovasc Medical, Inc., Or Yehuda, Israel) as a potential alternate therapy for patients with refractory angina who are not candidates for conventional revascularization procedures. Background: Increased coronary sinus (CS) pressure can reduce myocardial ischemia by redistribution of blood from nonischemic to ischemic territories. The Coronary Sinus Reducer is a percutaneous implantable device designed to establish CS narrowing and to elevate CS pressure. In preclinical experiments, implantation of the Reducer was safe and was associated with improved ischemic parameters. In the present study, the safety and feasibility of the Coronary Sinus Reducer was evaluated in patients with refractory angina who were not candidates for revascularization. Methods: Fifteen coronary artery disease patients with severe angina and reversible ischemia were electively treated with the Reducer. Clinical evaluation, dobutamine echocardiography, thallium single-photon emission computed tomography, and administration of an angina questionnaire were performed before and 6 months after implantation. Cardiac computed tomography was performed 2 days and 6 months after implantation. Results: All procedures were completed successfully. No procedure-related adverse events occurred during the periprocedural and the follow-up periods. Angina score improved in 12 of 14 patients. Average Canadian Cardiovascular Society score was 3.07 at baseline and 1.64 at follow-up (n = 14, p < 0.0001). Stress-induced ST-segment depression was reduced in 6 of 9 patients and was eliminated in 2 of these 6 (p = 0.047). The extent and severity of myocardial ischemia by dobutamine echocardiography and by thallium single-photon emission computed tomography was reduced (p = 0.004 [n = 13] and p = 0.042 [n = 10], respectively). Conclusions: Implantation of the Coronary Sinus Reducer is feasible and safe. These findings, along with the clinical improvement observed, support further evaluation of the Reducer as an alternative treatment for patients with chronic refractory angina who are not candidates for coronary revascularization.
UR - http://www.scopus.com/inward/record.url?scp=34247367219&partnerID=8YFLogxK
U2 - 10.1016/j.jacc.2007.01.061
DO - 10.1016/j.jacc.2007.01.061
M3 - ???researchoutput.researchoutputtypes.contributiontojournal.article???
C2 - 17466229
AN - SCOPUS:34247367219
SN - 0735-1097
VL - 49
SP - 1783
EP - 1789
JO - Journal of the American College of Cardiology
JF - Journal of the American College of Cardiology
IS - 17
ER -