Comparison between the Reflomat reflectance meter and the AutoAnalyzer for instant blood sugar measurements

M. Mimouni, E. Lew, L. Madanes, Z. Laron

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

The instrument tested in this study can be recommended for use in the clinic and at home. The coefficient of variation found in 20 replicate analyses of samples containing 100 to 200 mg/dl of glucose was found to be ±8.7 mg/dl for the Reflomat, and ±3.5 mg/dl for the Auto-Analyzer. Our findings show that the results obtained with the Reflomat generally compare well with those obtained with the AutoAnalyzer but they are unreliable in the range >300 mg/dl.

Original languageEnglish
Pages (from-to)990-991
Number of pages2
JournalIsrael Journal of Medical Sciences
Volume14
Issue number9
StatePublished - 1978
Externally publishedYes