Abstract
Aims Oesophageal temperature monitoring with single-sensor probe (SSP) has a variable ability to predict oesophageal ulceration as a consequence of pulmonary vein isolation (PVI). Multi-sensor self-expandable probes (MSPs) may offer improved thermal monitoring. The objective of this study was to compare the thermodynamic characteristics of both probes during PVI. Methods and results This prospective study enrolled 20 patients undergoing index PVI. Ten patients (group A) underwent dual monitoring with SSP and MSP and 10 control patients (group B) were monitored with SSP alone. Time to initial rise (>0.2°C), time to 1.0°C rise, peak temperature, and decay were recorded with each posterior wall lesion (20W, 198 applications). The operator was blinded to the MSP temperature data and ablation was only interrupted when SSP temperature increased by ≥2°C. All patients underwent endoscopy within 24 h. Initial temperature increase was detected earlier with MSP (13.4±7.5 vs. 30.5±15.4 s; P < 0.001); led to shorter time to 1.0°C rise (18.5±10.1 vs. 32.1±12.0 s; P < 0.001); and higher change in peak temperature (1.6±2.0 vs. 0.60±0.538C; P < 0.001). Decay time was similar between the probes (146.1±35.3 vs. 150.4±48.4 s; P = 0.89). The incidence of oesophageal ulceration was similar between the Groups A and B (5 and 4, respectively). Multi-sensor self-expandable probe provided greater sensitivity (100 vs. 60%) and similar specificity (60%) for detection of oesophageal ulceration. Five swine underwent oesophageal mapping before and after MSP placement without alteration in size or position. Conclusion Multi-sensor probes provide a superior thermodynamic profile. Its clinical value in reducing oesophageal injury requires further evaluation.
| Original language | English |
|---|---|
| Pages (from-to) | 891-897 |
| Number of pages | 7 |
| Journal | Europace |
| Volume | 17 |
| Issue number | 6 |
| DOIs | |
| State | Published - 1 Jun 2015 |
| Externally published | Yes |
Keywords
- Atrial fibrillation
- Multi-sensor probe
- Oesophageal injury
- Oesophageal temperature monitoring
- Pulmonary vein isolation
- Single-sensor probe