T1 - Comparing Effective Doses During Image-Guided Core Needle Biopsies with Computed Tomography Versus C-Arm Cone Beam CT Using Adult and Pediatric Phantoms
AU - Ben-Shlomo, A.
AU - Cohen, D.
AU - Bruckheimer, E.
AU - Bachar, G. N.
AU - Konstantinovsky, R.
AU - Birk, E.
AU - Atar, E.
N1 - Publisher Copyright:
© 2015, Springer Science+Business Media New York and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE).
PY - 2016/5/1
Y1 - 2016/5/1
N2 - Purpose: To compare the effective doses of needle biopsies based on dose measurements and simulations using adult and pediatric phantoms, between cone beam c-arm CT (CBCT) and CT. Method: Effective doses were calculated and compared based on measurements and Monte Carlo simulations of CT- and CBCT-guided biopsy procedures of the lungs, liver, and kidney using pediatric and adult phantoms. Results: The effective doses for pediatric and adult phantoms, using our standard protocols for upper, middle and lower lungs, liver, and kidney biopsies, were significantly lower under CBCT guidance than CT. The average effective dose for a 5-year old for these five biopsies was 0.36 ± 0.05 mSv with the standard CBCT exposure protocols and 2.13 ± 0.26 mSv with CT. The adult average effective dose for the five biopsies was 1.63 ± 0.22 mSv with the standard CBCT protocols and 8.22 ± 1.02 mSv using CT. The CT effective dose was higher than CBCT protocols for child and adult phantoms by 803 and 590 % for upper lung, 639 and 525 % for mid-lung, and 461 and 251 % for lower lung, respectively. Similarly, the effective dose was higher by 691 and 762 % for liver and 513 and 608 % for kidney biopsies. Conclusions: Based on measurements and simulations with pediatric and adult phantoms, radiation effective doses during image-guided needle biopsies of the lung, liver, and kidney are significantly lower with CBCT than with CT.
AB - Purpose: To compare the effective doses of needle biopsies based on dose measurements and simulations using adult and pediatric phantoms, between cone beam c-arm CT (CBCT) and CT. Method: Effective doses were calculated and compared based on measurements and Monte Carlo simulations of CT- and CBCT-guided biopsy procedures of the lungs, liver, and kidney using pediatric and adult phantoms. Results: The effective doses for pediatric and adult phantoms, using our standard protocols for upper, middle and lower lungs, liver, and kidney biopsies, were significantly lower under CBCT guidance than CT. The average effective dose for a 5-year old for these five biopsies was 0.36 ± 0.05 mSv with the standard CBCT exposure protocols and 2.13 ± 0.26 mSv with CT. The adult average effective dose for the five biopsies was 1.63 ± 0.22 mSv with the standard CBCT protocols and 8.22 ± 1.02 mSv using CT. The CT effective dose was higher than CBCT protocols for child and adult phantoms by 803 and 590 % for upper lung, 639 and 525 % for mid-lung, and 461 and 251 % for lower lung, respectively. Similarly, the effective dose was higher by 691 and 762 % for liver and 513 and 608 % for kidney biopsies. Conclusions: Based on measurements and simulations with pediatric and adult phantoms, radiation effective doses during image-guided needle biopsies of the lung, liver, and kidney are significantly lower with CBCT than with CT.
KW - Adult
KW - CT
KW - Cone beam CT
KW - Dosimetry
KW - Effective dose
KW - Image-guided needle biopsy
KW - Pediatric
KW - Radiation
KW - X-ray
UR - http://www.scopus.com/inward/record.url?scp=84944916661&partnerID=8YFLogxK
U2 - 10.1007/s00270-015-1227-8
DO - 10.1007/s00270-015-1227-8
M3 - ???researchoutput.researchoutputtypes.contributiontojournal.article???
C2 - 26493824
AN - SCOPUS:84944916661
VL - 39
SP - 732
EP - 739
JO - CardioVascular and Interventional Radiology
JF - CardioVascular and Interventional Radiology
SN - 7415-5101
IS - 5
ER -