TY - JOUR
T1 - Changes over time in the psycho-electric parameters in children with cochlear implants
AU - Henkin, Yael
AU - Kaplan-Neeman, Ricky
AU - Muchnik, Chava
AU - Kronenberg, Jona
AU - Hildesheimer, Minka
PY - 2003/7
Y1 - 2003/7
N2 - Information regarding typical changes in electrical stimulation levels during different stages post-implantation in children is scarce. Clinically, this information can serve as a general guideline to the nature and frequency of these changes, and consequently help to determine programming schedules that would optimize performance with the implant. Furthermore, it may have implications for the design of cochlear implants, and for monitoring the condition of the implanted cochlea. The main goal of the present study was, therefore, to evaluate changes over time in psycho-electric parameters, i.e. threshold levels, comfortable levels, and dynamic range, in children with the Nucleus 22 cochlear implant system. The maps of 37 prelingual children were examined at five time intervals: 1, 6, 12, 18 and 24 months post-connection. Maps were analyzed according to three cochlear segments: basal, medial, and apical. The results indicated significant changes in the psycho-electric parameters during the first 2 years post-implantation. Specifically, threshold (T) levels increased significantly during the first year and stabilized during the second year, whereas comfort (C) levels and dynamic range (DR) increased significantly during the entire 2-year follow-up. In addition, psycho-electric parameters varied among the different cochlear segments, resulting in a narrower DR in the basal one. Regardless of the underlying cause for these changes, the results of the present study suggest that, in children, the stabilization of psycho-electric parameters is a prolonged, gradual process. Since psycho-electric parameters play an important role in speech perception, these findings emphasize the need for frequent programming in the first 2 years of implant use in children.
AB - Information regarding typical changes in electrical stimulation levels during different stages post-implantation in children is scarce. Clinically, this information can serve as a general guideline to the nature and frequency of these changes, and consequently help to determine programming schedules that would optimize performance with the implant. Furthermore, it may have implications for the design of cochlear implants, and for monitoring the condition of the implanted cochlea. The main goal of the present study was, therefore, to evaluate changes over time in psycho-electric parameters, i.e. threshold levels, comfortable levels, and dynamic range, in children with the Nucleus 22 cochlear implant system. The maps of 37 prelingual children were examined at five time intervals: 1, 6, 12, 18 and 24 months post-connection. Maps were analyzed according to three cochlear segments: basal, medial, and apical. The results indicated significant changes in the psycho-electric parameters during the first 2 years post-implantation. Specifically, threshold (T) levels increased significantly during the first year and stabilized during the second year, whereas comfort (C) levels and dynamic range (DR) increased significantly during the entire 2-year follow-up. In addition, psycho-electric parameters varied among the different cochlear segments, resulting in a narrower DR in the basal one. Regardless of the underlying cause for these changes, the results of the present study suggest that, in children, the stabilization of psycho-electric parameters is a prolonged, gradual process. Since psycho-electric parameters play an important role in speech perception, these findings emphasize the need for frequent programming in the first 2 years of implant use in children.
KW - Children
KW - Cochlear implant
KW - Psycho-electric parameters
UR - http://www.scopus.com/inward/record.url?scp=0242475015&partnerID=8YFLogxK
U2 - 10.3109/14992020309078346
DO - 10.3109/14992020309078346
M3 - ???researchoutput.researchoutputtypes.contributiontojournal.article???
AN - SCOPUS:0242475015
SN - 1499-2027
VL - 42
SP - 274
EP - 278
JO - International Journal of Audiology
JF - International Journal of Audiology
IS - 5
ER -