A life-threatening arrhythmia induced by inappropriate activation of an implantable cardioverter defibrillator

Ehud Kadmon*, Jairo Kusniec, Boris Strasberg

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

We describe an unusual case of near fatal inappropriate implantable cardioverter defibrillator therapies due to atrial oversensing by a newly implanted ventricular lead. Chest X-ray revealed dislodgement of the active fixation lead to the tricuspid annulus area explaining the atrial oversensing and intermittent ventricular therapies.

Original languageEnglish
Pages (from-to)1716-1718
Number of pages3
JournalEuropace
Volume11
Issue number12
DOIs
StatePublished - Dec 2009

Fingerprint